AMI Codes-II O3 is a proven, low-maintenance monitor for the measurement of trace concentrations down to 1 ppb of ozone in water for pharmaceutical purposes. Using the robust colorimetric standard measurement method (DPD) according to DIN 38404-3, it is ideally suited to reliably prove the continuous protection of water storage systems against microbial growth as well as the absence of ozone before the water is used in a pharmaceutical production process.
Features
Your benefits
Keywords: Ozone analyzer, ozone sanitization, water for pharmaceutical purposes, purified water (PW), highly purified water (HPW), water for injection (WFI), water distribution and storage, on-line measurement
| Ozone measurement | |
| Measuring range | 0 to 500 ppb |
| Accuracy | ±1 ppb or 5% (whichever is greater) |
| Limit of detection | 1 ppb |
| Cycle time | 5 ... 10 min. (default 6') |
|
Transmitter Specifications |
|
| Power Supply | 100 ... 240 VAC, 50/60Hz or 24 VDC |
| Power consumption | max. 30 VA |
|
Sample conditions |
|
| Water consumption | min. 10 l/h |
| Water inlet pressure | 0.15 ... 2 bar |
| Sample temperature | 5 ... 50 °C |
|
Panel |
|
| Dimension | 400 x 850 x 200 mm |
| Material | stainless steel |
| Weight | 14.0 kg |