Colorimetric process analyzer (DPD-method) for measurement and dosing control of total chlorine and calculation of dichloramine.
Seperate chemical cleaning module available as accessory.

| Disinfectant measurement | Measuring range (Standard) [ppm] | Accuracy | |
| Total chlorine (tc1 & tc2) | 0.00 | 1.00 | ± 0.01 ppm |
| 1.00 | 3.00 | ± 0.06 ppm | |
| 3.00 | 5.00 | ± 0.2 ppm | |
| Measuring range (Extended) [ppm] | |||
| 0.00 | 10 | ± 10 % | |
| Dichloramine | Calculated difference of total chlorine 1 and 2. | ||
| Measurement time tc1 | 3 -5 sec. |
| Measurement time tc2 | 2 min. |
| Cycle time | 3 ... 60 min. |
| pH measuring range | 2 ... 12 |
| Temperature measuring range | -30 ... 100 °C |
|
Transmitter Specifications |
|
| Power Supply | 100 ... 240 VAC, 50/60Hz or 24 VDC |
| Power consumption | max. 30 VA |
|
Sample conditions |
|
| Water consumption | min. 10 l/h |
| Water inlet pressure | 0.15 ... 2 bar |
| Sample temperature | 5 ... 50 °C |
|
Panel |
|
| Dimension | 400 x 850 x 200 mm |
| Material | PVC |
| Weight | 9.0 kg |
Copyright © 2018-2019 swan-sh.com 瑞士swan水质分析仪 上海相泽机械 版权所有