Swansensor Shurecon P

——

打印本文             

  


Specifications

 

Recommended conductivity range 0.1 µS/cm ... 100 mS/cm
Accuracy ± 1.5% or ± 0.2 µS/cm; whichever is greater
Cell constant k ~0.4 cm-1
Temperature sensor type Pt1000, DIN class A

Operating conditions:
 
- normal operating temperature ≤ 50°C
- max. temperature short-time 90°C; at ambient pressure
- max. pressure 10 bar at 25°C

Materials:
 
- Shaft end tip PVDF
- Electrodes Platinum

Protection class

IP 68
Process connection on Ø16mm sensor head or Ø12mm shaft
Electrical connection Sensor with integrated cable with end sleeves
 

Description

Sensor for the measurement of specific conductivity.

For applications in surface water, potable water and cooling water.

For the use together with SWAN conductivity transmitters AMI Solicon4 or AMU Solicon4.